High-precision Thermal Mass Flow Meters & Mass Flow Controllers for Gases

Reliable technology and industry standard interfaces make the red-y smart series thermal mass flow meters (MFM) and mass flow controllers (MFC) particularly suitable for measurement and control in gas delivery systems and plant engineering applications.

PDF Datasheet

CORE FEATURES

√ Accurate Measurement

  • Accuracy up to ± 0.3% of full scale + ± 0.5% of reading
  • Turndown ratio 1 : 100

√ Analog & Digital

√ Operatiing Status Indication

√ Safe & Fast Control

√ get red-y Software

√ Options

  • Build-in display
  • Multigas
  • Profibus
  • Industrial Ethernet
  • Display and Control Device PCU-10
  • Gas Block Systems

√ 3-Years Warranty


Technical Specifications red-y smart series

MEASURING RANGES

Air/Full scale freely selectable

Body size ¼” from 0 … 25 mln/min to 0 … 60 ln/min (from 0 … 27 SCCM to 0 … 64 SLPM)
Body size ½” from 0 … 60 ln/min to 0 … 450 ln/min (from 0 … 65 SLPM to 0 … 480 SLPM)

INSTRUMENT VERSIONS

“Standard”
The economic solution

Accuracy: ± 1.0% of full scale*
Turndown ratio: 1 : 50

“Hi-Performance”
With highest accuracy and turndown ratio
(available for GSM < 200 ln/min (200 SLPM) /
GSC < 150 ln/min (150 SLPM) (air))

Accuracy: ± 0.3% of full scale + ± 0.5% of reading*
Turndown ratio: 1 : 100
*An additional error of ±0.25% may apply for analogue signals

PERFORMANCE DATA

Media (real gas calibration)

Air*, O2*, N2, He, Ar, CO2, H2, CH4, C3H8 (other gases and gas mixtures on request)
*O2 & N2 are calibrated with air

Response time

Meter (GSM): ± 80ms*; Controller (GSC): ± 500ms*
*depending on device configuration & according to SEMI standard E17-1011,
5-100% of range under optimized conditions

Repeatability

± 0.2% of full scale (according to SEMI standard E56-0309)

Longterm stability

< 1% of measured value / year

Power supply

24 Vdc (18 – 30 Vdc), 15 Vdc on request

Current consumption Standard

Meter (GSM): max. 100 mA; Controller (GSC): max. 250 mA
(GSC with valve type 8 max. 490mA)

Current consumption Profinet RT / EtherCAT

Meter (GSM): max. 125 mA; Controller (GSC): max. 340 mA
(GSC with valve type 8 max. 560mA)

Operation pressure

0.2 – 11 bar a (3 – 160 psia) / GSC with valve type 4.5 and 8 max. 8 bar a (120 psia)

Temperature (environment/gas)

0 – 50°C (32 – 122°F)

Body Materials

Anodized aluminum, optional stainless steel electropolished

Seals

FKM, EPDM, optional FFKM

Pressure sensitivity

< 0.2% / bar of reading (typical N2)

Temperature sensitivity

0.025% FS measuring range type per 1°C (<0.012% per 1°F)

Warm-up time

< 1 sec. for full accuracy

INTEGRATION & INSTALLATION

Output signals analog

0-20 mA, 4-20 mA, 0-5 V, 1-5 V, 0-10 V, 2-10 V

Output signals digital

RS-485; Modbus RTU (Slave); LabView-VIs available
Option: Profibus DP-V0, DP-V1 / Profinet RT / EtherCAT

Process connection

G¼” (BSPP* female) up to 60 ln/min (64 SLPM)
G½” (BSPP* female) up to 450 ln/min (480 SLPM)
*British Standard Pipe Parallel

Inlet section

None required

Electrical connection

Sub D plug, 9 pole

Mounting orientation

Any position (consult manufacturer above 5 bar or vertical mounting)

SAFETY

Test pressure

16 bara (240 psia)

Leak rate

< 1 x 10-6 mbar l/s He

Ingress protection class

IP50

EMC

EN 61326-1

 

 