High-precision Thermal Mass Flow Meters & Mass Flow Controllers for Gases
Reliable technology and industry standard interfaces make the red-y smart series thermal mass flow meters (MFM) and mass flow controllers (MFC) particularly suitable for measurement and control in gas delivery systems and plant engineering applications.
CORE FEATURES
√ Accurate Measurement
- Accuracy up to ± 0.3% of full scale + ± 0.5% of reading
- Turndown ratio 1 : 100
√ Analog & Digital
√ Operatiing Status Indication
√ Safe & Fast Control
√ get red-y Software
√ Options
- Build-in display
- Multigas
- Profibus
- Industrial Ethernet
- Display and Control Device PCU-10
- Gas Block Systems
√ 3-Years Warranty
Technical Specifications red-y smart series
MEASURING RANGES | |
---|---|
Air/Full scale freely selectable | Body size ¼” from 0 … 25 mln/min to 0 … 60 ln/min (from 0 … 27 SCCM to 0 … 64 SLPM) |
INSTRUMENT VERSIONS | |
“Standard” | Accuracy: ± 1.0% of full scale* |
“Hi-Performance” | Accuracy: ± 0.3% of full scale + ± 0.5% of reading* |
PERFORMANCE DATA | |
Media (real gas calibration) | Air*, O2*, N2, He, Ar, CO2, H2, CH4, C3H8 (other gases and gas mixtures on request) |
Response time | Meter (GSM): ± 80ms*; Controller (GSC): ± 500ms* |
Repeatability | ± 0.2% of full scale (according to SEMI standard E56-0309) |
Longterm stability | < 1% of measured value / year |
Power supply | 24 Vdc (18 – 30 Vdc), 15 Vdc on request |
Current consumption Standard | Meter (GSM): max. 100 mA; Controller (GSC): max. 250 mA |
Current consumption Profinet RT / EtherCAT | Meter (GSM): max. 125 mA; Controller (GSC): max. 340 mA |
Operation pressure | 0.2 – 11 bar a (3 – 160 psia) / GSC with valve type 4.5 and 8 max. 8 bar a (120 psia) |
Temperature (environment/gas) | 0 – 50°C (32 – 122°F) |
Body Materials | Anodized aluminum, optional stainless steel electropolished |
Seals | FKM, EPDM, optional FFKM |
Pressure sensitivity | < 0.2% / bar of reading (typical N2) |
Temperature sensitivity | 0.025% FS measuring range type per 1°C (<0.012% per 1°F) |
Warm-up time | < 1 sec. for full accuracy |
INTEGRATION & INSTALLATION | |
Output signals analog | 0-20 mA, 4-20 mA, 0-5 V, 1-5 V, 0-10 V, 2-10 V |
Output signals digital | RS-485; Modbus RTU (Slave); LabView-VIs available |
Process connection | G¼” (BSPP* female) up to 60 ln/min (64 SLPM) |
Inlet section | None required |
Electrical connection | Sub D plug, 9 pole |
Mounting orientation | Any position (consult manufacturer above 5 bar or vertical mounting) |
SAFETY | |
Test pressure | 16 bara (240 psia) |
Leak rate | < 1 x 10-6 mbar l/s He |
Ingress protection class | IP50 |
EMC | EN 61326-1 |